Touch Screen Display Automatic Manifold System

GM2-T
KIT-GMS-241M
Title:
Touch Screen Display Automatic Manifold System
GM2-T Series

GM2-T Series Touch Screen Display Medical Automatic Manifold System is designed to provide an uninterrupted gas supply without any manual adjustments. The system automatically switches over when th eprimary cylinder bank is depleted. Even in the case of a power failure, the system continues to supply gas without interruption. The system is designed to meet the latest edition of NFPA 99 and CGA standards.

Features

Automatic Changeover System

• Fully enclosed, dust-proof metal cabinet

• Automatic Switchover when pressure is below preset limit

• Touch Screen LCD Display for easy control and monitoring

• Automatically generated alarm table

• Built-in network connection, can be intergrated to the network system for real-time monitoring with RS-485 or ethernet cable

• Suitable for high flow systems; rated for 120m3/h (4200SCFH)* to 170m3/h (6000 SCFH)**

* When delivery pressure is 50 psi

** When delivery pressure is 180 psi

Pipeline

Silver brazing on piping joints for maximum leak prevention

System is designed to accommodate future expanison needs

Optional external filter provides easy replacement of filter element

Optional master shutoff valves

Headers have been tested to withstand high cylinder pressure

• Wall or floor mount available

GM2-TL- O2- U- ( 5L - 5R- S2 )
SeriesDelivery PressureGas ServiceStandard Code

Number of Cylinders

Manifold System Layout

Cylinder Valve Spacing
GM2-T

USA Standard   

L: 55 psi (380 MPa)

M: 100 psi (690 MPa) 

H: 185 psi (1270 MPa) 

 

EN Standard 

L: 72.5 psi (5 bar)

M: 116 psi (8 bar)

H: 145 psi (10 bar)

O2: Oxygen

AIR: Air

CO2: Carbon Dioxide

IN: Argon, Helium, Nitrogen

E: ISO 32

U: NFPA 99 (USA)

1L-2R: One cylinder on the left, Two cylinders on the right

5L-5R: Five cylinders on the left, Five cylinders on the Right

0-0: Left and Right side each with filter and master shutoff valve

Note: Direction of piping (Right or Left) is indicated by facing the manifold.

S: Standard layout

L: "L" Shape layout

U: "U" shape layout

D: Crossover layout

X: Staggered layout

1: 5" (127 mm)

2: 10" (254 mm)

3: 13" (330 mm)

4: 18" (457 mm)

Example: GM2-TL-O2-U-(5Lx5R-S2) indicates a 5*5 oxygen cylinder touch screen automatic manifold system. Distance between two cylinders is 10” on standard horizontal layout. GM2-TL-O2-U-(0x0) indicates an oxygen changeover system with filters and master shutoff valves. GM2-TL-O2-U indicates an oxygen changeover system only.

 

SeriesGas ServiceMax. Inlet Pressure psi (bar)Delivery Pressure psi (bar)Max. Delivery Flow SCFH (m3/h)Outlet ConnectionPigtail Specifications
GM2-TL-O2Oxygen3000 (207)10~145 (0.69~10)4200 (120)3/4" NPT attachment to the unionPigtail, CGA540
GM2-TM-INInert Gas3000 (207)10~230 (0.69~15.86)6000 (170)3/4" NPT attachment to the unionPigtail, CGA580
GM2-TM-CO2Carbon Dioxide2175 (150)4.4~125 (0.3~8.62)1060 (30)3/4" NPT attachment to the unionPigtail, CGA320
GM2-TH-AIRAir3000 (207)10~203 (0.69~14)5300 (150)3/4" NPT attachment to the unionPigtail, CGA346

* Connections can be changed to meet DIN, BS or other standards

 

• 24" or 36" flexible high pressure stainless steel braided pigtails with check valve

• Rigid copper pigtails are standard when gas service is oxygen

• Carbon Dioxide manifold systems are provided with H900G electric gas heater

• Siphon-type cylinder should not be used in the manifold system

Gas ServiceW in.(mm)H1 in.(mm)H2 in.(mm)
Oxygen, Air, Inert Gas41.3 (1050)15.8 (400)55.1 (1400)
Carbon Dioxide56.3 (1430)15.8 (400)55.1 (1400)

 

 

Standard Layout"L" Shape Layout"U"Shape LayoutCrossover LayoutStaggered Layout